Accuracy of the Velotron cycle ergometer and SRM Power Meter

Document Type

Journal Article

Faculty

Faculty of Computing, Health and Science

School

School of Exercise, Biomedical and Health Science / Centre for Exercise and Sports Science Research

Comments

Abbiss, C. R., Quod, M. J., Levin, G., Martin, D. T., & Laursen, P. B. (2009). Accuracy of the Velotron ergometer and SRM power meter. International journal of sports medicine, 30(02), 107-112. Available here

Abstract

The purpose of this study was to determine the accuracy of the Velotron cycle ergometer and the SRM power meter using a dynamic calibration rig over a range of exercise protocols commonly applied in laboratory settings. These trials included two sustained constant power trials (250 W and 414 W), two incremental power trials and three high-intensity interval power trials. To further compare the two systems, 15 subjects performed three dynamic 30 km performance time trials. The Velotron and SRM displayed accurate measurements of power during both constant power trials (

DOI

10.1055/s-0028-1103285

Share

 
COinS
 

Link to publisher version (DOI)

10.1055/s-0028-1103285